Peer Group Statistics Table 

Internal Quality Control Program (IQC) for Neonatal TSH Screening Test


TSH IQC NS MIS System [login]
中文


Supported by Preventive Medicine Foundation

Home Application Peer Group Statistics (Table 1) Month vs. Cumulative (Table 2) QC Chart of Individual Lab


Peer Group Statistics (Table 1)

Select LotNo:
Select Reagent Kit:
Monthly

Month :
   CI (Lot No.: TPC1901AE)  CII (Lot No.: TPC1901AF)
UnitID Reagent Kit
(Code)
Unit TargetMeann for
Mean
SDCV
(%)
TE
(%)
TEa
(%)
σ  TargetMeann for
Mean
SDCV
(%)
TE
(%)
TEa
(%)
σ
NS022 9 uU/mL blood 6.25.314 0.417.730.0200.7   27.625.014 2.4910.029.3201.1
Bias (%) = [ (│Mean - Target│) / Target ] x 100%
TE:Total Error(%) = Bias (%) + 2 × CV (%)
σ (Sigma) = [TEa% - Bias (%)] / CV (%)
[TOP]



Cumulative

Cumulative : from to
   CI (Lot No.: TPC1901AE)  CII (Lot No.: TPC1901AF)
UnitID Reagent Kit
(Code)
Unit TargetMeann for
Mean
SDCV
(%)
TE
(%)
TEa
(%)
σ  TargetMeann for
Mean
SDCV
(%)
TE
(%)
TEa
(%)
σ
NS002 9 uU/mL blood 6.1472 0.6911.322.6201.8   27.0472 2.8710.621.3201.9
NS022 9 uU/mL blood 6.25.6274 0.6411.432.5200.9   27.624.3274 2.4710.232.3200.8
NS022 11 uU/mL blood 6.36.5209 0.6510.023.2201.7   27.128.3209 2.498.822.0201.8
NS028 6 uU/mL blood 8.28.231 0.819.919.8202.0   31.132.231 2.297.117.8202.3
NS031 10 uU/mL blood 7.37.312 0.9713.326.6201.5   31.531.212 3.1910.221.4201.9
Bias (%) = [ (│Mean - Target│) / Target ] x 100%
TE:Total Error(%) = Bias (%) + 2 × CV (%)
σ (Sigma) = [TEa% - Bias (%)] / CV (%)
[TOP]

Reagent Kit Reagent Code
Guangzhou Fenghua Bionegineering CO. LTD.6
PerkinElmer-AutoDELFIA11
PerkinElmer-DELFIA10
PerkinElmer-GSP9
Laboratory prepared99



Home Application Peer Group Statistics (Table 1) Month vs. Cumulative (Table 2) QC Chart of Individual Lab

Contact us
Address: P.O. Box 26-624 Taipei, Taiwan, 10699  Tel:+886-2-703-6080
    e-mail:
Fax: +886-2-2703-6070

Supported by:
      Preventive Medicine Foundation, Taipei, Taiwan
          updated : 27 Feb, 2020