各檢驗單位 TSH 篩檢檢驗 內部品管統計結果

各檢驗單位 TSH 篩檢檢驗 內部品管統計結果(總表一)

選擇 LotNo:
選擇 Reagent Kit:
月統計表

統計月份:
   CI (Lot No.: TPC1502AC)  CII (Lot No.: TPC1502AD)
UnitID Reagent Kit
(Code)
Unit TargetMeann for
Mean
SDCV
(%)
TE
(%)
TEa
(%)
σ  TargetMeann for
Mean
SDCV
(%)
TE
(%)
TEa
(%)
σ
NS002 11 uU/mL blood 8.17.421 0.7510.128.9100.1   36.137.821 3.228.521.7100.6
NS004 9 uU/mL blood 6.36.421 0.6410.021.6100.8   26.228.521 3.2011.231.2100.1
NS007 9 uU/mL blood 6.36.664 0.649.724.2252.1   26.728.364 2.9510.426.8251.8
NS022 11 uU/mL blood 6.87.01 ---30-   30.023.51 ---30-
誤差 (Bias)% = { [ │累計平均值 (Mean) - 目標值 (Target)│] / 目標值 (Target) } x 100%
總誤差 (TE, Total Error)% = 誤差 (Bias)% + 2 × 累計變異係數 (CV)
σ (Sigma) = (TEa% - 誤差 (Bias)%) / 累計變異係數 (CV)
[TOP]



累計統計表

累計期間:自
   CI (Lot No.: TPC1502AC)  CII (Lot No.: TPC1502AD)
UnitID Reagent Kit
(Code)
Unit TargetMeann for
Mean
SDCV
(%)
TE
(%)
TEa
(%)
σ  TargetMeann for
Mean
SDCV
(%)
TE
(%)
TEa
(%)
σ
NS001 11 uU/mL blood 7.76.312 0.548.635.310-1.0   31.931.112 2.678.619.7100.9
NS002 11 uU/mL blood 8.18.146 1.5218.837.5100.5   36.137.246 3.268.820.6100.8
NS004 9 uU/mL blood 6.36.4396 0.6410.021.6100.8   26.226.9396 3.0611.425.4100.6
NS007 9 uU/mL blood 6.36.41491 1.8528.959.4250.8   26.727.21491 3.2812.126.0251.9
NS022 11 uU/mL blood 6.87.5271 0.7610.130.6302.0   30.029.8271 3.2310.822.3302.7
誤差 (Bias)% = { [ │累計平均值 (Mean) - 目標值 (Target)│] / 目標值 (Target) } x 100%
總誤差 (TE, Total Error)% = 誤差 (Bias)% + 2 × 累計變異係數 (CV)
σ (Sigma) = (TEa% - 誤差 (Bias)%) / 累計變異係數 (CV)
[TOP]

Reagent Kit Reagent Code
PerkinElmer-AutoDELFIA11
PerkinElmer-GSP9
Laboratory prepared99