Internal Quality Control Program (IQC) for Neonatal TSH Screening Test
TSH IQC NS MIS System [login]
Supported by Preventive Medicine Foundation
| CI | CII |
QC Control Lot No. | TPC2101AG | TPC2101AH |
ReagentKit | PerkinElmer-DELFIA |
Unit | uU/mL blood |
Duration of the Analyzing | Month (2024/10) | CUM (2024/08/01~2024/10/31) | Month (2024/10) | CUM (2024/08/01~2024/10/31) |
Runs (N) | 12 | 18 | 12 | 18 |
Mean | 7.6 | 7.7 | 42.2 | 41.5 |
SD | 1.24 | 1.45 | 3.15 | 3.53 |
CV (%) | 16.3 | 18.8 | 7.5 | 8.5 |
Target Value | 8.4 | 8.4 | 41.5 | 41.5 |
Total Error (%) | 42.2 | 46.0 | 16.6 | 17.0 |
TEa (%) | 20 | 20 | 20 | 20 |
σ | 0.6 | 0.6 | 2.4 | 2.4 |
Bias (%) = [ (│Mean - Target│) / Target ] x 100%
TE:Total Error(%) = Bias (%) + 2 × CV (%)
σ (Sigma) = [TEa% - Bias (%)] / CV (%)
|
Lot No.: TPC2101AG ; Reagent : PerkinElmer-DELFIA ; Unit: uU/mL blood
Duration : 2024-08-01 to 2024-10-31 ; Target : 8.4 ; SD : 0.84 ( ●) | |
Lot No.: TPC2101AH ; Reagent : PerkinElmer-DELFIA ; Unit: uU/mL blood
Duration : 2024-08-01 to 2024-10-31 ; Target : 41.5 ; SD : 4.85 ( □) | |
 |
Lot No.: TPC2101AG ; Reagent : PerkinElmer-DELFIA ; Unit : uU/mL blood
Duration : 2024-08-01 to 2024-10-31 ; Target : 8.4 ; SD : 0.84 | |
 |
Lot No.: TPC2101AH ; Reagent : PerkinElmer-DELFIA ; Unit : uU/mL blood
Duration : 2024-08-01 to 2024-10-31 ; Target : 41.5 ; SD : 4.85 | |
 |
Contact us
Address:
P.O. Box 26-624 Taipei, Taiwan, 10699 Tel:+886-2-703-6080
e-mail:
Fax:
+886-2-2703-6070
Supported by:
Preventive Medicine Foundation, Taipei, Taiwan
updated
: 29 Mar, 2022