Internal Quality Control Program (IQC) for Neonatal TSH Screening Test
TSH IQC NS MIS System [login]
Supported by Preventive Medicine Foundation
| CI | CII |
QC Control Lot No. | TPC1901AE | TPC1901AF |
ReagentKit | PerkinElmer-DELFIA |
Unit | uU/mL blood |
Duration of the Analyzing | Month (2020/10) | CUM (2020/09/01~2020/10/31) | Month (2020/10) | CUM (2020/09/01~2020/10/31) |
Runs (N) | 10 | 12 | 10 | 12 |
Mean | 7.6 | 7.3 | 31.4 | 31.2 |
SD | 0.80 | 0.97 | 3.43 | 3.19 |
CV (%) | 10.5 | 13.3 | 10.9 | 10.2 |
Target Value | 7.3 | 7.3 | 31.5 | 31.5 |
Total Error (%) | 25.2 | 26.6 | 22.2 | 21.4 |
TEa (%) | 20 | 20 | 20 | 20 |
σ | 1.5 | 1.5 | 1.8 | 1.9 |
Bias (%) = [ (│Mean - Target│) / Target ] x 100%
TE:Total Error(%) = Bias (%) + 2 × CV (%)
σ (Sigma) = [TEa% - Bias (%)] / CV (%)
|
Lot No.: TPC1901AE ; Reagent : PerkinElmer-DELFIA ; Unit: uU/mL blood
Duration : 2020-09-01 to 2020-10-31 ; Target : 7.3 ; SD : 0.92 ( ●) | |
Lot No.: TPC1901AF ; Reagent : PerkinElmer-DELFIA ; Unit: uU/mL blood
Duration : 2020-09-01 to 2020-10-31 ; Target : 31.5 ; SD : 3.33 ( □) | |
 |
Lot No.: TPC1901AE ; Reagent : PerkinElmer-DELFIA ; Unit : uU/mL blood
Duration : 2020-09-01 to 2020-10-31 ; Target : 7.3 ; SD : 0.92 | |
 |
Lot No.: TPC1901AF ; Reagent : PerkinElmer-DELFIA ; Unit : uU/mL blood
Duration : 2020-09-01 to 2020-10-31 ; Target : 31.5 ; SD : 3.33 | |
 |
Contact us
Address:
P.O. Box 26-624 Taipei, Taiwan, 10699 Tel:+886-2-703-6080
e-mail:
Fax:
+886-2-2703-6070
Supported by:
Preventive Medicine Foundation, Taipei, Taiwan
updated
: 29 Mar, 2022