Internal Quality Control Program (IQC) for Neonatal TSH Screening Test
TSH IQC NS MIS System [login]
Supported by Preventive Medicine Foundation
| | CI | CII |
| QC Control Lot No. | TPC1502AC | TPC1502AD |
| ReagentKit | PerkinElmer-AutoDELFIA |
| Unit | uU/mL blood |
| Duration of the Analyzing | Month (2020/06) | CUM (2018/07/01~2020/06/30) | Month (2020/06) | CUM (2018/07/01~2020/06/30) |
| Runs (N) | 1 | 271 | 1 | 271 |
| Mean | 7.0 | 7.5 | 23.5 | 29.8 |
| SD | - | 0.76 | - | 3.23 |
| CV (%) | - | 10.1 | - | 10.8 |
| Target Value | 6.8 | 6.8 | 30.0 | 30.0 |
| Total Error (%) | - | 30.6 | - | 22.3 |
| TEa (%) | 30 | 30 | 30 | 30 |
| σ | - | 2.0 | - | 2.7 |
Bias (%) = [ (│Mean - Target│) / Target ] x 100%
TE:Total Error(%) = Bias (%) + 2 × CV (%)
σ (Sigma) = [TEa% - Bias (%)] / CV (%)
|
Lot No.: TPC1502AC ; Reagent : PerkinElmer-AutoDELFIA ; Unit: uU/mL blood
Duration : 2018-07-01 to 2020-06-30 ; Target : 6.8 ; SD : 0.58 ( ●) | |
Lot No.: TPC1502AD ; Reagent : PerkinElmer-AutoDELFIA ; Unit: uU/mL blood
Duration : 2018-07-01 to 2020-06-30 ; Target : 30.0 ; SD : 3.80 ( □) | |
 |
Lot No.: TPC1502AC ; Reagent : PerkinElmer-AutoDELFIA ; Unit : uU/mL blood
Duration : 2018-07-01 to 2020-06-30 ; Target : 6.8 ; SD : 0.58 | |
 |
Lot No.: TPC1502AD ; Reagent : PerkinElmer-AutoDELFIA ; Unit : uU/mL blood
Duration : 2018-07-01 to 2020-06-30 ; Target : 30.0 ; SD : 3.80 | |
 |
Contact us
Address:
P.O. Box 26-624 Taipei, Taiwan, 10699 Tel:+886-2-703-6080
e-mail:
Fax:
+886-2-2703-6070
Supported by:
Preventive Medicine Foundation, Taipei, Taiwan
updated
: 29 Mar, 2022