Internal Quality Control Program (IQC) for Neonatal TSH Screening Test
TSH IQC NS MIS System [login]
Supported by Preventive Medicine Foundation
| CI | CII |
QC Control Lot No. | TPC1502AC | TPC1502AD |
ReagentKit | PerkinElmer-AutoDELFIA |
Unit | uU/mL blood |
Duration of the Analyzing | Month (2020/01) | CUM (2020/01/01~2020/01/31) | Month (2020/01) | CUM (2020/01/01~2020/01/31) |
Runs (N) | 12 | 12 | 12 | 12 |
Mean | 6.3 | 6.3 | 31.1 | 31.1 |
SD | 0.54 | 0.54 | 2.67 | 2.67 |
CV (%) | 8.6 | 8.6 | 8.6 | 8.6 |
Target Value | 7.7 | 7.7 | 31.9 | 31.9 |
Total Error (%) | 35.3 | 35.3 | 19.7 | 19.7 |
TEa (%) | 10 | 10 | 10 | 10 |
σ | -1.0 | -1.0 | 0.9 | 0.9 |
Bias (%) = [ (│Mean - Target│) / Target ] x 100%
TE:Total Error(%) = Bias (%) + 2 × CV (%)
σ (Sigma) = [TEa% - Bias (%)] / CV (%)
|
Lot No.: TPC1502AC ; Reagent : PerkinElmer-AutoDELFIA ; Unit: uU/mL blood
Duration : 2020-01-01 to 2020-01-31 ; Target : 7.7 ; SD : 0.47 ( ●) | |
Lot No.: TPC1502AD ; Reagent : PerkinElmer-AutoDELFIA ; Unit: uU/mL blood
Duration : 2020-01-01 to 2020-01-31 ; Target : 31.9 ; SD : 2.00 ( □) | |
 |
Lot No.: TPC1502AC ; Reagent : PerkinElmer-AutoDELFIA ; Unit : uU/mL blood
Duration : 2020-01-01 to 2020-01-31 ; Target : 7.7 ; SD : 0.47 | |
 |
Lot No.: TPC1502AD ; Reagent : PerkinElmer-AutoDELFIA ; Unit : uU/mL blood
Duration : 2020-01-01 to 2020-01-31 ; Target : 31.9 ; SD : 2.00 | |
 |
Contact us
Address:
P.O. Box 26-624 Taipei, Taiwan, 10699 Tel:+886-2-703-6080
e-mail:
Fax:
+886-2-2703-6070
Supported by:
Preventive Medicine Foundation, Taipei, Taiwan
updated