Peer Group Statistics Table 

Internal Quality Control Program (IQC) for Neonatal TSH Screening Test


TSH IQC NS MIS System [login]
中文


Supported by Preventive Medicine Foundation

Home Application Peer Group Statistics (Table 1) Month vs. Cumulative (Table 2) QC Chart of Individual Lab


Peer Group Statistics (Table 1)

Select LotNo:
Select Reagent Kit:
Monthly

Month :
   CI (Lot No.: TPC1502AC)  CII (Lot No.: TPC1502AD)
UnitID Reagent Kit
(Code)
Unit TargetMeann for
Mean
SDCV
(%)
TE
(%)
TEa
(%)
σ  TargetMeann for
Mean
SDCV
(%)
TE
(%)
TEa
(%)
σ
NS004 9 uU/mL blood 6.06.415 0.477.321.4100.5   25.226.515 2.077.820.8100.6
Bias (%) = [ (│Mean - Target│) / Target ] x 100%
TE:Total Error(%) = Bias (%) + 2 × CV (%)
σ (Sigma) = [TEa% - Bias (%)] / CV (%)
[TOP]



Cumulative

Cumulative : from to
   CI (Lot No.: TPC1502AC)  CII (Lot No.: TPC1502AD)
UnitID Reagent Kit
(Code)
Unit TargetMeann for
Mean
SDCV
(%)
TE
(%)
TEa
(%)
σ  TargetMeann for
Mean
SDCV
(%)
TE
(%)
TEa
(%)
σ
NS001 11 uU/mL blood 7.76.312 0.548.635.310-1.0   31.931.112 2.678.619.7100.9
NS002 9 uU/mL blood 8.16.396 0.619.741.610-1.3   36.127.296 2.659.744.110-1.5
NS004 9 uU/mL blood 6.06.3975 0.6310.025.0100.5   25.227.0975 2.8010.427.9100.3
NS007 9 uU/mL blood 6.36.52619 1.5123.249.6250.9   26.727.82619 3.5112.629.4251.7
NS022 11 uU/mL blood 6.87.5271 0.7610.130.6302.0   30.029.8271 3.2310.822.3302.7
Bias (%) = [ (│Mean - Target│) / Target ] x 100%
TE:Total Error(%) = Bias (%) + 2 × CV (%)
σ (Sigma) = [TEa% - Bias (%)] / CV (%)
[TOP]

Reagent Kit Reagent Code
PerkinElmer-AutoDELFIA11
PerkinElmer-GSP9
Laboratory prepared99



Home Application Peer Group Statistics (Table 1) Month vs. Cumulative (Table 2) QC Chart of Individual Lab

Contact us
Address: P.O. Box 26-624 Taipei, Taiwan, 10699  Tel:+886-2-703-6080
    e-mail:
Fax: +886-2-2703-6070

Supported by:
      Preventive Medicine Foundation, Taipei, Taiwan
          updated : 27 Feb, 2020

[ Home ]